Where this LCD applies
Each contract number is a jurisdiction on the remittance; the policy binds claims processed under these contracts and no others.
Billing and coding: diagnoses and procedure codes
Since 2019 the codes live in the companion article rather than the LCD. Billing and Coding A56915 (Billing and Coding: Urine Drug Testing) carries the diagnosis and procedure lists the contractor loads as the claims edit. CPT codes are shown as bare numbers because the descriptors are licensed by the AMA; HCPCS Level II descriptors are public and shown.
A56915: Billing and Coding: Urine Drug Testing (Billing and Coding, effective 2026-06-01)
- Covered ICD-10-CM codes
- 424
- 1 group
- Non-covered ICD-10-CM codes
- 0
- Procedure codes listed
- 66
- Full article
- cms.gov record
| ICD-10-CM | Description (FY2027) |
|---|---|
| E87.20 | — |
| E87.21 | — |
| E87.22 | — |
| E87.29 | — |
| F11.11 | — |
| F11.120 | — |
| F11.13 | — |
| F11.20 | — |
| F11.220 | — |
| F11.221 | — |
| F11.222 | — |
| F11.229 | — |
| F11.23 | — |
| F11.24 | — |
| F11.250 | — |
| F11.251 | — |
| F11.259 | — |
| F11.281 | — |
| F11.282 | — |
| F11.288 | — |
| F11.29 | — |
| F12.10 | — |
| F12.11 | — |
| F12.13 | — |
Procedure codes: 80305, 80306, 80307, 80320, 80321, 80322, 80323, 80324, 80325, 80326, 80327, 80328, 80329, 80330, 80331, 80332, 80333, 80334, 80335, 80336, 80337, 80338, 80339, 80340, 80341, 80342, 80343, 80344, 80345, 80346, 80347, 80348, 80349, 80350, 80351, 80352, 80353, 80354, 80355, 80356 and 26 more in the article.
Coverage indications, limitations and medical necessity
Purpose
Urine drug testing (UDT) provides timely, objective, and actionable information to clinicians by identifying the presence or absence of drugs of potential abuse in the body to assist the clinician in making treatment decisions. 1
This policy details:
• The appropriate indications and allowed number of UDTs billed over time for safe medication management of prescribed substances in risk stratified pain management patients and/or in identifying and treating substance use disorders (SUDs);
• Designates documentation, by the clinician caring for the beneficiary in the beneficiary’s medical record, of medical necessity for, and testing ordered on an individual patient basis;
• Provides an overview of presumptive UDT and definitive UDT testing by various methodologies.
This policy addresses UDT for Medicare patients only.
Definitions
As used in this document, the following terminology relates to the basic forms of UDT:
• Presumptive/Qualitative Drug Testing (hereafter called "presumptive" UDT) - Covered when medically necessary to immediately determine the presence or absence of drugs or drug classes in a urine sample; results expressed as negative or positive or as a numerical result; includes competitive immunoassays (IA) and thin layer chromatography. 2
• Definitive/Quantitative/Confirmation (hereafter called “definitive” UDT) - Covered when clinically indicated and medically reasonable and necessary based on this LCD to identify specific medications, illicit substances, and metabolites; reports the results of analytes absent or present typically in concentrations such as ng/mL; definitive methods include but are not limited to GC-MS and LC-MS/MS testing methods only. 2
• Specimen Validity Testing - Urine specimen testing to ensure that it is consistent with normal human urine and has not been adulterated or substituted, may include, but is not limited to pH, specific gravity, oxidants, and creatinine. This however is quality assurance, not a Medicare benefit, and thus not separately payable by Medicare.
• Immunoassay (IA) - Ordered by clinicians primarily to identify the presence or absence of drug classes and some specific drugs; biochemical tests that measure the presence above a cutoff level of a substance (drug) with the use of an antibody; read by photometric technology. 2
• Point of Care Testing (POCT) - Covered when medically necessary by clinicians caring for the beneficiary for immediate test results for the immediate management of the beneficiary; available when the beneficiary and physician are in the same location; IA test method that primarily identifies drug classes and a few specific drugs; platform consists of cups, dipsticks, cassettes, or strips; read by the human eye, or read by instrument assisted direct optical observation. 3
• Standing Orders - Test request for a specific patient representing repetitive testing to monitor a condition or disease for a limited number of sequential visits; individualized orders for certain patients for pre-determined tests based on historical use, risk, and community trend patient profiles; clinician can alter the standing order. Note: A profile is developed based on specific characteristics of a specific patient, while a panel is a general non-specific group of tests that may have unnecessary tests for the specific patient being treated.
• Blanket Orders - Test request that is not for a specific patient; rather, it is an identical order for all patients in a clinician’s practice without individualized decision making at every visit.
• Reflex Testing - Laboratory testing that is performed "reflexively" after initial test results to identify further diagnostic information essential to patient care. This testing is not necessarily based on a specific physician's order.
Drug Testing Methods
The Clinical Laboratory Improvement Amendments (CLIA) regulates laboratory testing and requires clinical labs to be certified by their State as well as the Center for Medicare & Medicaid Services (CMS) before they can accept human samples for diagnostic testing. 3 Multiple types of CLIA certificates may be obtained based on the complexity of testing a lab conducts. CLIA levels of complexity (CLIA-waived, moderate complexity, and high complexity) are addressed only as they correspond to the Healthcare Common Procedure Coding System (HCPCS) code description in the related billing and coding article.
A. Presumptive Testing Methods
• Presumptive UDT:
A presumptive UDT that consists of various platforms including cards, dipsticks, cassettes, and cups based on qualitative competitive immunoassay methodology with one or more analytes in the test. A presumptive IA test detects the presence of the amount of drug/substance present in urine above a predetermined “cut-off” value and may be read by direct optical observation or by instrument assisted direct optical observation.
A positive test result is reported when the concentration of drug is above the cutoff; a negative is reported when the concentration of drug is below the cut-off. Positive test results are presumptive but not necessarily definitive due to sensitivity and cross-reactivity limitations. 4 Negative test results do not necessarily indicate the absence of a drug or substance in the urine specimen. 3 The accuracy of the results of a presumptive UDT will depend on the testing environment, type of test, the drug being tested for, and training of the individual conducting the test. This type of test should only be used when results are needed immediately.
Presumptive UDT by Instrumented Chemistry Analyzers:
Chemistry analyzers with IA UDT technology can be used in an office or clinical laboratory setting. This test provides less immediate test results. At no time is IA technology by chemistry analyzer analysis considered confirmatory (definitive) testing.
A presumptive positive IA test detects the presence of a drug/substance in urine at or above the “cut-off” value. If the concentration of the drug is below the cut-off, the result will be negative. Presumptive positive tests are not always true positives due to sensitivity, specificity, and cross-reactivity limitations. Negative test results do not necessarily indicate the absence of a drug or substance in the urine specimen. 3
Food and Drug Administration (FDA) approved/cleared test platforms are available in the marketplace as well as laboratory developed tests (LDTs) such as modified FDA approved/ cleared and non-FDA approved/cleared platforms and/or reagents. LDTs generally have been modified to test at a lower cutoff in order to detect substances that would have been missed at a higher cutoff. For example, an FDA labeled cutoff may be 300 ng/mL and the LDT cutoff for the same drug may be 100 ng/mL. 3
Presumptive UDT can be carried out at any validated cut-off concentration. Lowering of the cut-off concentration provides more stringent cutoffs for illicit drugs. LDTs may include non-FDA cleared tests not available in CLIA-waived or moderate complexity tests (e.g., tramadol, tapentadol, carisoprodol, fentanyl, zolpidem). Lowering the cutoff increases the possibility of detecting a drug when the test has been modified from the recipe of the manufacturer.
Limitations of Presumptive UDT:
Presumptive UDT testing is limited due to:
• Primarily screens for drug classes rather than specific drugs, and therefore, the practitioner may not be able to determine if a different drug within the same class is causing the positive result;
• Produces erroneous results due to cross-reactivity with other compounds or does not detect all drugs within a drug class; 5
• Given that not all prescription medications or synthetic/analog drugs are detectable and/or have assays available, it is unclear as to whether other drugs are present when some tests are reported as positive;
• Cut-off may be too high to detect presence of a drug. 5
These limitations may mean that presumptive testing is insufficient for certain clinical needs.
An IA involves an antibody that reacts best with the stimulating drug and reacts to a lesser extent (cross-reactive) or not at all with other drugs in the drug class. While presumptive tests vary in their ability to detect illicit drugs such as tetrahydrocannabinol (THC), cocaine, 3,4-methylenedioxy-N-methylamphetamine (MDMA; “ecstasy”), and phencyclidine (PCP), they may not be optimal tests for many prescription drugs, such as: opiates, barbiturates, benzodiazepines, and opioids.
For example, opiate reagents are formulated from morphine. Consequently, the cross-reactivity for other opioids and opiates varies based on the manufacturer and lot number. The semisynthetic opioids, hydromorphone and hydrocodone, may contribute to a positive presumptive result, while the semisynthetic opioids, oxycodone and oxymorphone, will not typically be detected even at a 300 ng/mL cutoff. Synthetic opioids, such as fentanyl, meperidine, and methadone, will not be detected by current opiate IA testing. Consequently, a positive opiate result by IA normally necessitates more specific identification of the substance(s) that account for the positive result, and a negative result does not rule out the presence of opiates or opioids. 6
Presumptive UDT reagents for benzodiazepine are typically formulated for oxazepam, a metabolite of diazepam (Valium ® ) and chlordiazepoxide (Librium ® ), the main benzodiazepines prescribed 20 years ago. However, many of the more than 10 benzodiazepines that are currently available do not cross-react with IA benzodiazepine reagents. In particular, clonazepam and lorazepam give false negative results with presumptive IA tests and may necessitate more specific identification to account for the negative result. Similarly, a positive screening test result may require definitive UDT to identify the specific drug(s).
Synthetic/analog or “designer” drugs manufactured to elude law enforcement require definitive testing for detection. Most commercially available IA reagents fail to detect designer drugs such as psychedelic phenethylamines even at very high concentrations.
In summary, presumptive IA UDT is often unable to identify specific drugs within many drug classes, particularly within the amphetamine, barbiturate, benzodiazepine, tricyclic antidepressants, and opiate/opioid drug classes. Drugs such as buprenorphine, amphetamines, benzodiazepines, and cocaine/heroin may yield false negative IA results due to low cross-reactivity or non-reactivity, and drugs such as fentanyl, carisoprodol, tramadol, tapentadol, and synthetic designer drugs cannot be detected by presumptive IA. Therefore, it may be medically necessary for clinicians to utilize definitive UDT when the presumptive tests for these drugs are negative. 1,5
B. Definitive UDT:
Gas Chromatography coupled with Mass Spectrometry (GC-MS) and Liquid Chromatography coupled with Mass Spectrometry (LC-MS/MS) are complex technologies that use the separation capabilities of gaseous or liquid chromatography with the analytical capabilities of mass spectrometry. These methodologies require the competency of on-site highly trained experts in this technology and interpretation of results. While these tests require different sample preparation and analytical runs, they identify specific drugs, metabolites, and most illicit substances and report the results as absent or present typically in concentrations of ng/mL. 2
Quantification should not be used to determine adherence with a specific dosage or time of dose of a pain medication or illicit drug for clinical purposes. Rather, the use of quantitative drug data may be important for many reasons such as in a differential patient assessment. 1 For example, when several opioids are present in the urine of a patient prescribed a single opioid, quantification may help the clinician decide whether the presence of the other opioids is consistent with metabolism of the prescribed opioid, opioid contamination during manufacturing, or if more than 1 drug within a class is being used.
Quantification may also provide information in the setting of illicit drug use. Serial creatinine-corrected quantitative values may assist in the differential assessment of ongoing drug use or cessation of drug use with continued drug excretion. 1
• GC-MS
GC-MS can only be performed on molecules that are volatile. If the test drug is not volatile in its own right, it must be modified or derivatized to a volatile form. To derivatize, the test drug must be extracted from the urine, eluted from the extraction device, concentrated, and then reacted with a chemical reagent to make a volatile product. Each drug class may require a different derivatizing agent. For patients on multiple classes of medications, laboratories using GC procedures must make different volatile derivatives in order to perform comprehensive testing. Since a GC column may not be able to separate more than one class of compounds, multiple chromatographic runs on different column types may be required to monitor multiple drug classes 1 . Newer GC-MS instruments often use tandem systems. GC-MS methodology allows for the testing of multiple substances but differs in ease of run.
• LC-MS/MS
LC-MS/MS is roughly 100 times more sensitive and selective, involves fewer human steps, provides quicker turn-around time, uses less specimen volume, and can test for a larger number of substances simultaneously when compared to GC-MS. 1 After sample preparation, it is injected into the LC-MS/MS. The sample has to undergo hydrolysis to break the glucuronide bond that frees the drug and drug metabolites. Hydrolysis is followed by multiple additional steps including protein precipitation, centrifugation, and purification. Deuterium-labeled isotopic internal standards are added to quantify the drugs and drug metabolites.
The sample is injected when the mobile phase is flowing through the chromatographic column. Each drug and drug metabolite interacts with the mobile phase and stationary phase differently and moves at different speeds depending on their chemical properties. In other words, each analyte elutes at different times. Specific drugs and metabolites are identified by their retention time and quantified against isotopic internal standards for each drug and metabolite. Each drug peak has a minimum of 2 mass transmissions to be compared to drug standards (calibrators) to ensure identification.
CLIA-Certified Laboratories-Informational only
CLIA specifies quality standards for proficiency testing, facility administration, general laboratory systems, pre-analytic, analytic, and post-analytic systems, onsite supervision requirements, personnel qualifications and responsibilities, quality control, and quality assessment. 3
High complexity laboratories must ensure that testing is carried out by onsite qualified, trained personnel using validated reliable methods compliant with regulatory procedures (42 CFR Part 493). Both GC-MS and LC-MS/MS require a quality program to monitor the quality and audit the competency of the staff. LC-MS/MS instrument maintenance must be performed daily as well as the validation of instrument performance prior to patient specimens. Final review and approval of GC-MS and LC-MS/MS results must be performed by a qualified clinical laboratory scientist as defined in 42 CFR Part 493.1489 (Testing Personnel Qualifications). A GC-MS or LC-MS/MS laboratory must have a qualified laboratory director, qualified physician, or qualified clinical laboratory scientist, as provided in 42 CFR 493.1443 (Laboratory Director Qualifications).
Purpose of UDT:
Presumptive UDT may be ordered by the clinician caring for a beneficiary when it is necessary to rapidly obtain and/or integrate results into clinical assessment and treatment decisions.
Definitive UDT is considered reasonable and necessary when the clinical information supplied supports the definitive testing as in:
• Identify a specific substance or metabolite that is inadequately detected by a presumptive UDT screen;
• Definitively identify specific drugs in a large family of drugs;
• Identify a specific substance or metabolite that is not detected by presumptive UDT such as fentanyl, meperidine, synthetic cannabinoids, and other synthetic/analog drugs;
• Identify drugs when a definitive concentration of a drug is needed to guide management (e.g., discontinuation of THC use according to a treatment plan);
• Identify a negative, or confirm a positive, presumptive UDT result that is inconsistent with a patient’s self-report, presentation, medical history, or current prescribed pain medication plan;
• Rule out an error as the cause of a presumptive UDT result;
• Identify non-prescribed medication or illicit use for ongoing safe prescribing of controlled substances; and
• Use in a differential assessment of medication efficacy, side effects, or drug-drug interactions.
Definitive UDT may be reasonable and necessary based on patient specific indications, including historical use, medication response, and clinical assessment, when accurate results are necessary to make clinical decisions. 1 To establish that a test is reasonable and necessary, the clinician’s rationale for the definitive UDT and the tests ordered must be documented in the patient’s medical record.
Drug Testing Panels
• Presumptive UDT Panels
Presumptive UDT typically involves testing for multiple analytes based on the specific beneficiary's clinical history and risk assessment and must be documented in the medical record. May be ordered as a panel and billed a "Per Patient encounter" regardless of the number of analytes tested.
• Definitive UDT Panels
Physician-directed definitive profile testing is reasonable and necessary when ordered for a particular patient based upon historical use, clinical findings, and community trends . However, the same physician-defined profile is not reasonable and necessary for every patient in a physician’s practice. Definitive UDT orders should be individualized based on clinical history and risk assessment and must be documented in the medical record.
Specimen Type
Urine or oral fluid is the preferred biologic specimen for testing because of the ease of collection, storage, and cost-effectiveness. 1 UDT cannot detect the dosage of drug ingested/used, the time of use, or the means of delivery (intravenous vs. oral vs. inhaled). Detection time of a substance in urine is typically 1-3 days depending on the drug, rate of metabolism, and rate of excretion. Lipid-soluble drugs, such as marijuana, may remain in body fat and be detected upwards of a week or more.
Ethanol is not discussed in this LCD:
Note: Ethanol is a known drug of abuse but is routinely tested in blood, not urine. In addition, the DEA Resource Guide 7 states that alcohol is exempt from control by the Controlled Substances Act (CSA).
Covered Indications for UDT
Group A – Symptomatic patients, Multiple drug ingestion, and/or Patients with unreliable history
A patient who presents in a variety of medical settings with signs or symptoms of substance use toxicity will be treated presumptively to stabilize the patient while awaiting presumptive, then definitive testing to determine the cause(s) of the presentation. The need for definitive UDT is based upon presumptive test findings, responses to medical interventions, and treatment plan. 1 A presumptive UDT should be performed as part of the evaluation and management of a patient who presents in an emergency room or urgent care setting with any 1 of the following:
• Coma;
• Altered mental status in the absence of a clinically defined toxic syndrome or toxidrome;
• Severe or unexplained cardiovascular instability (cardiotoxicity);
• Unexplained metabolic or respiratory acidosis in the absence of a clinically defined toxic syndrome or toxidrome;
• Seizures with an undetermined history;
• To provide antagonist to specific drug.
The presumptive findings, definitive drug tests ordered, and reasons for the testing must be documented in the patient’s medical record.
Group B - Diagnosis and treatment for substance abuse or dependence
A patient in active treatment for a SUD or monitoring across different phases of recovery may undergo medical management for a variety of medical conditions. A physician who is writing prescriptions for medications to treat either the SUD or other conditions may need to know if the patient is taking substances which can interact with prescribed medications or taking prescribed medications as expected. 1 The risk of drug-drug interactions is inherent to the patient and may be compounded by prescribed medications. UDT is a medically necessary and useful component of chemical dependency diagnosis and treatment. The UDT result influences treatment and level of care decisions. 3 Ordered tests and testing methods (presumptive and/or definitive) must match the stage of screening, treatment, or recovery; the documented history; and Diagnostic and Statistical Manual of Mental Disorders (DSM V) diagnosis. For patients with no known indicators of risk for SUD, the clinician may screen for a broad range of commonly abused drugs using presumptive UDT. For patients with known indicators of risk for SUD, the clinician may screen for a broad range of commonly abused drugs using definitive UDT. For patients with a diagnosed SUD, the clinician should perform random UDT at random intervals to properly monitor the patient 3 . Testing profiles must be determined by the clinician based on the following medical necessity guidance criteria:
• Patient history, physical examination, and previous laboratory findings;
• Stage of treatment or recovery;
• Suspected abused substance;
• Substances that may present high risk for additive or synergistic interactions with prescribed medication (e.g., benzodiazepines, alcohol).
The patient’s medical record must include an appropriate number of UDTs billed over time based on the stage of screening, treatment, or recovery; 8 and the rationale for the drugs/drug classes ordered; the results must be documented in the medical record and used to direct care. 3
The policy text continues in the CMS record.
Summary of evidence (opening)
Providers should consider the prescribed medications, risk assessment, and clinical presentation when selecting which tests to order.
The contractor cites 16 sources in the bibliography; the full summary and analysis of evidence are in the CMS record.
Dates, lineage and related policies
- Original determination effective
- 2015-10-01
- Current revision effective
- 2025-08-28
- Last reviewed by the contractor
- 2025-07-18
- MCD version
- 51
- Derived from
- L32450
The contractor lists 2 National Coverage Determinations as related: NCD 130.5 Treatment of Alcoholism and Drug Abuse in a Freestanding Clinic, NCD 130.6 Treatment of Drug Abuse (Chemical Dependency). Where an NCD speaks, it controls; the LCD can only address what the NCD leaves open.
Other related documents: A59562 (Response to Comments).
Using this policy on a claim
Match the documented indication to the covered indications above before the service is scheduled, carry a diagnosis from the article's covered list on the claim line, and keep the elements the documentation section asks for in the record, because the contractor can request it later through medical review. A denial under this policy arrives as CARC 50 with remark N115; the LCD lookup guide walks through the appeal path and the Advance Beneficiary Notice rules, and the Wisconsin Physicians Service Insurance Corporation hub lists every other active policy from the same contractor.
The same policy title at other contractors
Contractors often adopt each other's policies and then revise them separately, so the criteria and the diagnosis lists drift apart. The topic comparison lines up every version.
Frequently asked questions
What does LCD L34645 cover?
• Specimen Validity Testing - Urine specimen testing to ensure that it is consistent with normal human urine and has not been adulterated or substituted, may include, but is not limited to pH, specific gravity, oxidants, and creatinine. This however is quality assurance, not a Medicare benefit, and thus not separately payable by Medicare. The full indications and limitations are reproduced on this page from the CMS Medicare Coverage Database export of September 24, 2026.
Which states does LCD L34645 apply to?
Wisconsin Physicians Service Insurance Corporation applies it to Medicare claims in AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. A Local Coverage Determination binds only the contractor that wrote it; the same service in another jurisdiction is judged under that contractor's own policy or, where none exists, claim by claim.
Which diagnosis codes support medical necessity under LCD L34645?
The companion billing and coding article A56915 lists 424 ICD-10-CM codes in 1 group that support medical necessity; the first 24 appear on this page and the complete list is in the article on cms.gov.
How do I appeal a denial under LCD L34645?
The remittance carries claim adjustment reason code 50 with remark code N115, naming the LCD. Compare the documented indication with the policy's covered indications and the article's diagnosis list, then file a redetermination within 120 days with the record attached; if the service genuinely falls outside the policy, the patient can be billed only when a valid Advance Beneficiary Notice was obtained before the service.
Sources
Every figure on this page is taken from the CMS publications below, as released by the Centers for Medicare & Medicaid Services. Projection built 2026-10-02. Verify against the primary file before billing or contracting decisions.
- Medicare Coverage Database, current LCD exportVersion MCD release 2026-09-24 · effective 2026-09-20 · file lcd.csvSHA-256 2fcc4251b6ddd1eb…
- Medicare Coverage Database, current Billing and Coding Articles exportVersion MCD release 2026-09-24 · effective 2026-09-20 · file article.csvSHA-256 f31932f1df3b4035…
- ICD-10-CM FY2027 code descriptionsVersion FY2027 · effective 2026-10-01 · file icd10cm_codes_2027.txtSHA-256 3c0583a38ee0e848…
Disclaimer
The policy text and code lists are reproduced from the CMS Medicare Coverage Database export as an operational reference. Verify against the current LCD and article on cms.gov before billing; coverage depends on the full record and the contractor. Not legal, clinical or billing advice.